      Insulation work of medical device shell

      Release time:2023-03-30

      Medical diagnosis, measurement and treatment equipment using AC power supply may cause leakage current due to improper grounding and electrical insulation, which may expose patients and even medical staff to the risk of electric shock, burns, internal organ damage and arrhythmia. So do a good jobMedical device shellInsulation work.

      美容儀器殼9.jpg


      1. The equipment shall be manufactured and sealed to prevent contact with live parts and parts that may be live in a single fault state. If the bulb can be replaced without tools, it must be ensured to prevent contact with the live parts of the lamp when loading and unloading the bulb.


      2. The position or size of any hole on the top cover of the medical device shell shall ensure that the test rod with a diameter of 4 mm and a length of 100 mm will not touch live parts when it is freely and vertically suspended in the access hole throughout its length.


      3. When the handle, knob, control lever, etc. are removed, the conductor part of the operating mechanism of the control device can be touched. When the open circuit test voltage is not greater than 50V AC and the test current is not less than 1a, the resistance measured to the protective grounding terminal should not be greater than 0.2 Ω; Or it should be isolated from live parts by adopting a method of 17g in GB 9706.1-2007.


      4. If the components with line voltage above 25V AC or 60V DC in the shell of the medical device cannot be disconnected from the power supply through the external main switch or the plug device that can be pulled out at any time, additional covers shall be used for protection, so that contact can be prevented even after the shell is opened, or when the spaces are separated and arranged, the "live" mark shall be clearly marked.


      5. In order to prevent contact with live parts, the shell must be removed only with tools, otherwise an automatic device should be used when opening or removing the shell, so that these parts are not used.


      6. When the device has a preset controller, it can be adjusted by the user with tools during normal use. The hole opened for adjusting these control devices shall be designed so that the adjusting tool cannot contact the basic insulation or any live parts in the hole, or any parts isolated from the grid power supply part only through the basic insulation but without grounding protection.


      Copyright: Shenzhen Zhongyuan Plastic Co., Ltd. | mold processing customized manufacturer. If you want to know the processing price of the factory, how much is it, and which is good, please contact us Yue ICP Bei No. 17060480
      主站蜘蛛池模板: 综合国产精品第一页| 久久综合丝袜日本网| 国产精品综合色区在线观看 | 亚洲国产综合专区在线电影| 亚洲伊人成无码综合网 | 日韩亚洲国产综合高清| 伊人久久亚洲综合影院| 国产成人亚洲综合色影视| 亚洲国产成人久久综合碰| 国产色婷婷五月精品综合在线| 亚洲国产成人久久综合碰| 欧美色综合久久久久久| 国产综合精品一区二区三区| 亚洲狠狠成人综合网| 涩涩色中文综合亚洲| 热综合一本伊人久久精品| 狠狠综合久久AV一区二区三区| 国产亚洲精品精品国产亚洲综合| 欧美亚洲综合色在| 亚洲伊人久久综合影院| 亚洲伊人成无码综合网| 国产成+人+综合+欧美亚洲| 欧美日韩国产综合一区二区三区| 一本色道久久88综合日韩精品 | 狠狠色综合网站久久久久久久高清 | 国产色婷婷五月精品综合在线| 99久久综合国产精品二区| 色综合.com| 国产精品天干天干综合网| 婷婷亚洲综合五月天小说| 久久综合狠狠综合久久激情 | 天堂无码久久综合东京热| 国产亚洲精品精品国产亚洲综合 | 色综合合久久天天综合绕视看| 午夜激情影院综合| 涩涩色中文综合亚洲| 激情综合亚洲色婷婷五月| 欧美综合自拍亚洲综合图片区| 亚洲欧美精品综合中文字幕| 六月婷婷缴清综合在线| 狠狠色丁香婷婷综合|